BHAST  Biased Highly Accelerated Stress Test Chamber

BHAST Biased Highly Accelerated Stress Test Chamber

BHAST Chamber This method is used to evaluate the operational reliability of non-hermetically sealed IC devices (solid-state components) under high temperature and humidity conditions, and performs accelerated stress aging tests under temperature, humidity, and bias conditions for chips, semiconductors, and other electronic components.

Description

Test standard

AEC Q101 (Automotive Grade IC)
JIS C0096-2
GB/T2423.40-1997 Environmental Testing of Electrical and Electronic Products - Part 2: Test Methods - Section Cx: Constant Humid Heat (Unsaturated High-Pressure Steam)
IEC60068-2-66-1994 Environmental Testing - Part 2-66: Test Methods - Test Cx: Steady-State Humid Heat
JESD22-A100 Temperature and Humidity Bias Life Test (Cyclic)
JESD22-A101 Steady-State Temperature, Humidity/Bias Life Test (Temperature, Humidity, Bias Life Test)
JESD22-A102 High Pressure Steam Test (Accelerated Moisture Resistance)
JESD22-A108 Temperature, Bias Voltage, and Operating Life
JESD22-A110 HAST (Highly Accelerated Temperature Humidity Stress Test)

Application areas:

PCB, LCD board, battery, capacitor, resistor, IC semiconductor, connector, circuit board, magnetic material, polymer material, EVA, photovoltaic module, automotive-grade chip

Product features:

1. The inner chamber features a double-layer, curved design, preventing condensation and dripping, thus ensuring that the product is not subjected to direct impact from superheated steam during testing, which could affect the results.
2. An efficient vacuum pump maintains optimal pure saturated steam conditions inside the chamber.
3. Equipped with a 7-inch full-color touchscreen, 250 programmable cycles with 12,500 segments each, USB data download function, and an RS-485 communication interface.
4. Automotive-grade silicone sealing strip for excellent airtightness and durability.
5. Automatic water replenishment function with a front-mounted water level indicator.
6. Uses a dry-bulb/wet-bulb sensor for direct measurement (control modes include: dry-bulb, unsaturated, and saturated humidity).

BHAST  Biased Highly Accelerated Stress Test Chamber
Model WBE-HAST-30 WBE-HAST-40 WBE-HAST-55
Volume (L) 31L 68L 154L
Internal Dimensions
(
×D) (mm)
350×450mm 400×550mm 550x650mm
Power 2.8KW 3.2KW 4KW
Weight 252KG 285KG 358KG
Temperature Range
Temperature Fluctuation ≤±0.5℃
Temperature Uniformity ≤±0.5℃
Humidity Range 65% ~ 100% RH
Humidity Fluctuation ≤±2 %R.H
Humidity Uniformity ≤±3 % R.H
Pressure Range 0.2~3kg/cm² (0.018~0.294Mpa);0.2~4kg/cm² (0.018~0.394Mpa)
Heating Time Ambient temp. to +155℃: Approx. 45 min
Pressurization Time Atmospheric to 3 kg/cm²: Approx. 35 min (with boiler);
Approx. 5 min (with external gas)
Pressure Deviation ≤±2Kpa
Air Circulation Centrifugal fan + air duct circulation system
Pressurization Method 1. Boiler steam pressurization; 2. External gas pressurization
Resolution Temp.: ±0.1℃, Hum.: ±0.1%RH, Pressure: 0.01 kg/cm², Time: 0.01 min
Insulation & Structure Interior: SUS316 Stainless Steel; Exterior: Cold-rolled steel with electrostatic coating; Insulation: Superfine glass wool
Noise ≤60 dB