Explosion-proof High and Low Temperature Test Chamber: Also known as a high and low temperature explosion-proof chamber, it is a specially designed environmental simulation testing device. It is used to test the performance of products or materials under extreme temperature conditions (high temperature, low temperature, temperature cycling, temperature change), and also has explosion-proof safety features, making it suitable for industrial environments with flammable and explosive risks.
Product features
1. Intelligent gas monitoring and emergency response system: Real-time dynamic monitoring of flammable and explosive gas concentration changes.
2. Explosion-proof safety design: The front door adopts a secondary explosion-proof design, equipped with an explosion-proof lock and explosion-proof chain.
3. Automatic fire extinguishing system: Built-in smoke sensor automatically activates the carbon dioxide fire extinguishing device when spontaneous combustion is detected.
4. Scientifically designed air ducts: Rapid and uniform temperature and humidity control.
5. Designed to accommodate both special temperature and humidity points and commonly used experimental points, resulting in more energy-efficient equipment operation.
6. Teflon-coated inner casing: High temperature resistance, corrosion resistance, and flame retardancy.
WBE Explosion-proof Climate Test Chamber Models&Specification
![]()
Explosion-proof High and Low Temperature Climate Test Chamber Specification
【💡:Learn More Chamber】
| Explosion-proof High and Low Temperature Climate Test Chamber |
| Performance |
Temp. Range |
A:-40~+150℃,B:-20~+150℃,D:-70~+150℃(+180℃) |
| Temp.Fluctuation |
≤±0.3℃ |
| Temp.Uniformity |
≤1.5℃ |
| Temp.Deviation |
≤±1℃ |
| Heating rate |
≥4℃/min, average throughout, approximately 55 minutes from -70℃ to +150℃. |
| Cooling rate |
≥1℃/min, average throughout, +150→-70℃ takes approximately 115 minutes. |
| Humidity Range |
20%RH~98%RH |
| Humidity Uniformity |
≤3.0%RH |
| Humidity Fluctuation |
≤±2.0%RH |
Safety Protection Function |
Safety protection |
Pressure relief, explosion-proof windows, explosion-proof chains (standard configuration), forced ventilation (optional) |
| Monitoring function |
Smoke detection, gas detection (CH, CO, H2), over-temperature protection |
Fire Extinguishing Function |
Perfluorohexanone, water spray,water flooding |
| Safety PLC |
Manual/automatic fire suppression, remote fire suppression, independent alarmand power cut-off |
Chamber Material |
Inner Chamber |
SUS#304 Stainless Steel |
| Outer Chamber |
Cold-rolled Steel (double-sided electrostatic spraying treatment) |
| Cooling Method |
Air-cooled |